User Manual 301/302
FREQuency:RASTer:SOURce EXTernal | INTernal
Purpose
This command selects the source of the sample clock generator. This command affects both the standard
and the arbitrary waveforms.
<EXTernal> selects an external source
<INTernal> selects the internal source
FREQuency:RASTer:SOURce?
Response
The 302 will return EXT if an external source is selected, or INT if the internal source is selected.
FUNCTion:MODE FIXed | USER | SEQuence
Purpose
This command defines the type of waveform that will be available at the output connector.
<FIXed> selects standard waveforms
<USER> selects arbitrary waveforms
<SEQuence> selects sequenced waveforms
FUNCTion:MODE?
Response
The 302 will return FIX, USER, or SEQ depending on the present 302 setting.
FUNCtion:SHAPe SINusoid | TRIangle | SQUare | PULSe | RAMP | SINC | EXPonential |
GAUSsian | DC}
Purpose
This command defines the shape of waveform that will be available at the output connector. This command
has no affect if present 302 setting is USER or SEQ.
<SINusoid> selects sine waveform
<TRIangle> selects triangular waveform
<SQUare> selects square waveform
<PULSe> selects pulse waveform
<RAMP> selects ramp waveform
<SINC> selects sinc waveform
<EXPonential> selects exponential waveform
<GAUSsian selects gaussian waveform
<DC> selects dc waveform
FUNCtion:SHAPe?
Response
The 302 will return SIN, TRI, SQU, PULS, RAMP, SINC, EXP, GAUS, or DC depending on the present 302