MEASURING PROCEDURES
General Procedures:
❶
When connecting or disconnecting test leads to
a circuit, always first turn off power to device or circuit being tested and
discharge all capacitors.
❷
If magnitude of signal is not known, set
switch to highest range and reduce until satisfactory reading is
obtained.
❸
Strictly observe the max input limits.
D • Meßprozeduren
Allgemein: ❶ Vor Verbinden und Trennen der Meßkabel mit dem Schaltkreis,
diesen abschalten und Kondensatoren entladen, ❷ Bei unbekannter Signalgröße,
bei höchstem Bereich beginnen and dann niedriger schalten bis gate Auflösung
erreicht wird. ❸ Maximale Grenzen nicht überschreiten.
E • Procedimientos de medida
En general:. ❶ Cuando vaya a aplicar o retirar las pumas de prueba a/de un cir-
cuito, en primer lugar desconecte siempre la alimentación del dispositivo o cir-
cuito sometido a prueba y descargue todos los condensadores. ❷ si no conoce
la magnitud de la señal, ponga el selector en la escala más alta y vaya reducien-
do hasta obtener una lectura satisfactoria, ❸ Observe estrictamente los límites
máximos de entrada .
F • Procédures de Mesure
Général: ❶ Avant de connecter ou de déconnecter les cordons de test, coupez I
lálimentation du circuit mesuré et déchargez les condensateurs. ❷ Si la magni-
tude du signal n’est pas connue, commencez avec le calibre le plus élevé, et
diminuez ensuite jusqu'à obtenir une bonne lecture. ❸ Ne dépassez pas les lim-
ites d'entrée.
DC AND VOLTAGE MEASUREMENT (See Fig 1)
❶
Connect red test lead to V-W Input and black test lead to COM Input.
❷
Set
Function/Range Switch to desired DCV or ACV position.
❸
Touch Probe tips
across voltage source (in parallel with circuit).
❹
Voltage value will appear on
Digital Display along with the voltage polarity (negative DC measurements).
NNoottee ffoorr AACC MMeeaassuurreemmeennttss::
The AC converter used in the HD110, HD110T and
320 models is "average-responding." An RMS value is determined by first recti-
fying and filtering the signal to obtain the average value. The average value is
then scaled upward by a factor of 1.11 to obtain the RMS value. Different scaling
factors need to be applied for other waveforms (1 for square waves; 1.733 for
triangle; 3.247 for full wave rectified sine wave; 2.591 for half wave rectified sine
wave). Please note that these correction factors only apply for standard wave-
forms that are free from noise and distortion. True RMS values of nonsinusoidal
– 8 –